ELECTROSURGERY AND ENERGISED #
%%{init: {'theme':'dark'}}%% mindmap root((Electrosurgery)) Defi Electrosurgery High fz AC - heat Cut coagulate Electrocautery DC - heat Biophy electric current Types Monopolar Bipolar
Introduction #
Definitions #
- Electrosurgery:
- Uses high-frequency alternating current (AC) to generate heat,
- which is used to cut or coagulate tissue.
- Electrocautery:
- Uses direct current (DC) to generate heat at the tip of an instrument.
Monopolar #
Active Electrode:
- a small, precise electrode that generates heat at its tip
- that delivers the high-frequency electrical current to the tissue.
- allowing for cutting or coagulating tissue at the point of contact.
Return Electrode:
- aka dispersive electrode or grounding pad
- The larger electrode placed elsewhere on the patientβs body.
- which safely return the electrical current from the patient back to the electrosurgical unit (ESU).
- Because of its larger surface area, the current is dispersed over a wide area, minimizing heat production.
Biophysics #
What is Electrical Current? #
- Electrical Current: The flow of electrons through tissue between adjacent atoms.
- Heat Generation:
- As the current flows through tissue, the tissue resists the flow,
- converting the electrical energy into heat.
Principle of Diathermy #
- Electrosurgical Unit (ESU): Generates the electrical current used in electrosurgery.
- Heat Generation:
- Depends on current density, which is the amount of current applied per ==unit area==.
- Maximum Heat:
- Generated where the current density is highest,
- typically at the tip of the active electrode.
- Return Electrode:
- In monopolar circuits, a larger electrode placed in body,
- dissipating the current and minimizing heat at that site.
- Temperature Rise is proportional to the time the active electrode is in contact with the tissue.
Frequency of Electrical Current #
- High-Frequency Current:
- Used by the ESU, ranging from 200 kHz to 3.3 MHz.
- Low frequency current:
- Used by household appliances
- Using standard AC frequency (60 Hz) would cause unwanted neuromuscular stimulation.
Current Waveforms in Electrosurgery #
1. Cutting Waveform (Low Voltage Continuous Current) #
- Wave: Continuous, sine wave
- Time: ESU on 100% of the time.
- Voltage: low
- Effects:
- Cutting:
- When the active electrode is held slightly away from the tissue,
- it quickly vaporizes the tissue.
- Desiccation:
- When the active electrode touches the tissue directly,
- it dries out the cells, forming a coagulum.
- Cutting:
2. Coagulation Waveform (High Voltage Intermittent Current) #
- Wave: Intermittent
- Time: The ESU is on only 6% of the time and off 94%.
- Voltage: High.
- Effects:
- Coagulation:
- Causes more heat to spread through the tissue,
- leading to
- protein denaturation,
- tissue shrinking, and
- forming a coagulum.
- Fulguration (Spray):
- Creates electrical arcing in coagulation mode that chars tissue over a wide area.
- Active electrode is held slightly away the tissue
- uses the highest voltage setting, so caution is needed.
- Coagulation:
3. Blended Waveform #
- Wave: Combination of both coagulation and cutting waveform.
- Time: The ESU is on 50% of the time and off 50%.
- Voltage: Moderate.
- Effects:
- Provides a mix of cutting and coagulation, with varying levels of each.
Waveform Type | Time (ESU Output) | Voltage | Effects |
---|---|---|---|
Cutting Waveform | Continuous (100% on) | Low | - Cutting: Vaporizes tissue when the active electrode is slightly away. |
- Desiccation: Dries out tissue to form a coagulum when the active electrode touches the tissue directly. | |||
Coagulation Waveform | Intermittent (6% on, 94% off) | High | - Coagulation: Causes protein denaturation, tissue shrinking, and forms a coagulum due to thermal spread. |
- Fulguration (Spray): Electrical arcing chars tissue when the active electrode is slightly away. | |||
Blended Waveform | 50% on, 50% off | Moderate | - Produces a combination of cutting and coagulation, with varying degrees of each. |
Electrosurgical Circuit Types #
1. Monopolar Delivery System #
- Circuit Flow: ESU β Active Electrode β Target Tissue β Patient’s Body β Return Electrode β ESU
flowchart LR A[ESU] --> B[Active Electrode] --> C[Target Tissue] --> D[Patient's Body] --> E[Return Electrode] -->|Current Flow| A
Key Points for Placing the Return Electrode:
- Material should be made of low-resistant material.
- Surface Area should be large.
- Placement Location:
- Should be placed over well-vascularized muscle mass, avoiding crossing a joint to maximize conduction of current.
- Ideally placed in the same quadrant as the operative site to ensure the current travels the shortest distance through the patient’s tissues.
- Placement Condition: Ensure the electrode is not wrinkled.
- Avoid:
- Vascular insufficient areas.
- High resistance areas (e.g., scar tissue, edematous tissue, bony prominences).
Preventing Alternative Current Pathways:
- Ensure metal prostheses and ECG electrodes are out of the circuit’s direct path.
- Remove all metal piercings.
- The patient should not be touching any metal part of the table.
Newer ESUs:
- Feature Contact Quality Monitors (CQM) that shut down the unit if the contact surface area becomes too small.
2. Bipolar Delivery System #
Circuit Flow: ESU β Active Tine β Target Tissue (between tines) β Return Tine β ESU**
Key Features:
- Patient’s body does not form part of the circuit.
- Can be safely used on narrow tissue pedicles (e.g., penis).
- Monopolar use in such cases may cause high current densities in the pedicle, leading to thermal injury.
Safety Guidelines for Electrosurgical Units (ESU) #
1. Training and usage #
- Train all team members on the proper use of the ESU.
- Use the lowest effective power setting to minimize risk.
2. Implants #
Pacemaker Considerations #
- Avoid using diathermy with patients who have pacemakers.
- If diathermy is necessary:
- Prefer bipolar diathermy.
- If monopolar diathermy must be used:
- Use short bursts.
- Place the return electrode as far away as possible from the pacemaker.
- Intraoperative
- Ensure continuous ECG monitoring during surgery.
- Have a defibrillator readily available in the theatre.
Warning
Pacemaker settings #
- Set the pacemaker to “maintain only” mode to prevent inappropriate shock delivery due to electrical interference.
- In Emergency Surgery If pacemaker settings cannot be adjusted, place a magnet over the ICD to inhibit shock delivery.
Cochlear Implants #
- Bipolar diathermy should be used.
- If monopolar diathermy is necessary:
- Place the return electrode as close as possible to the active electrode.
3. Skin Preparation #
- Be cautious with spirit-based fluids as they can ignite and cause serious burns.
- Avoid pooling of fluids in areas such as
- the umbilicus,
- under the drapes, or
- around the return electrode.
- Allow adequate time for the fluid to dry before proceeding.
- Avoid pooling of fluids in areas such as
Electrosurgery in Endoscopy #
1. Monopolar Snare Polypectomy #
- Pedunculated Polyps:
- Use blended mode.
- Right Colon Polyps:
- Exercise extreme caution due to the thinner wall, which increases the risk of perforation.
2. Sphincterotomy in ERCP #
- Active electrode: sphincterotome has a monofilament or braided wire.
- Ensure short duration of contact time with tissue to minimize damage.
3. Argon Plasma Coagulation (APC) #
- APC Overview:
- A form of non-contact monopolar electrosurgery used for coagulation and fulguration.
- Key Differences from Conventional Fulguration:
- Utilizes a stream of inert argon gas passing over the tip, which gets ionized, confining the current to pass through the gas and onto the tissue.
- Allows precision application of current to tissues.
Question
How argon plasma coagulation, coagulate a large area?
- The current follows the path of least resistance, thus avoiding already coagulated tissue, thus rapidly coagulating a larger area.
- Uses:
- Endoscopy Setting:
- Coagulation of superficial vascular structures (e.g., angiodysplasia).
- Palliative ablation of tumors.
- Hemostasis following the resection of large polyps.
- Operative Setting:
- Hepatic resection.
- Retinal surgery.
- Endoscopy Setting:
Tissue Energizers: Advanced Bipolar and Ultrasonic Devices #
1. Advanced Bipolar Devices #
Electrothermal Bipolar Vessel Sealing (EBVS) Devices #
- Examples: Ligasure, Enseal
- How they work:
- Combine bipolar current with mechanical compression
- Can seal blood vessels up to 7mm in diameter
- Process:
- High frequency, low voltage current denatures vessel wall proteins, elastin and collagen
- Mechanical pressure forms a coagulum (seal)
- Sensors in tip measure tissue impedance to ensure complete seal
- Energy delivery stops automatically when seal is complete
- Tissue is then cut mechanically
2. Ultrasonic Devices #
Harmonic Scalpel #
- Converts electrical energy to ultrahigh frequency mechanical energy (55,500 Hz/second)
- How it works:
- Mechanical energy generates heat
- Heat denatures proteins, forming a coagulum to seal vessels
- Can seal vessels up to 5-7mm
- Maximum temperature: 150Β°C (minimizes thermal spread)
- Cutting mechanisms:
- Heat-induced cellular vaporization ("cavitational cutting and fragmentation")
- Tissue stretching beyond elastic limit
LOTUS (LaparOscopic dissection by Torsional UltraSound) #
- Uses torsional ultrasound to minimize “distal drilling effect”
3. Combined Energy Devices #
Thunderbeat #
- Uses both ultrasonic and bipolar energy
- Two modes:
- Seal and cut: simultaneous ultrasonic and bipolar energy
- Seal: purely bipolar energy for vessel sealing
- Sensors measure tissue impedance to prevent excess energy delivery
Note: Choice of device depends on preference, availability, and cost. All of above devices are almost equal in clinical trials.
Hazards of Electrosurgery #
A. Electrothermal Injuries #
General Surgery Risks #
- Diathermy pad burns
- Insulation failure
- Breaks in active electrode insulation (common in distal third)
- Current can discharge from defects
Laparoscopic Surgery Specific Risks #
- Off-camera injuries (1-5 per 1000 laparoscopies)
Direct Coupling #
- If the electrosurgical unit (ESU) is accidentally activated while the active electrode is close to another metal instrument, current can flow through that second instrument.
Capacitance Coupling #
A capacitor forms when two conductors are separated by a non-conductor.
An electrostatic field is created between the conductors, allowing alternating current (AC) to pass through (but not direct current).
How It Works: The active electrode (Conductor 1) can pass current through intact insulation (Non-conductor) into an adjacent metal object (Conductor 2) like a trocar or even nearby bowel.
Prevention: Activate electrosurgical device only after the electrode is in contact with the target tissue, the energy will go directly into the tissue.
Danger
- Don’t activate monopolar diathermy before the electrode touches the target tissue.
- the current might pass through the insulation into the adjacent conductor.
Warning
- Capacitance coupling in open surgery
- A towel clip can cause similar issues if used to hold the electrosurgery wires.
- If the towel clip touches the patient, it could cause burns.
B. Fire and Explosion Risks #
- Caused by alcohol-based skin preparations
C. Electrosurgical Smoke #
- Carcinogenic
- Irritates respiratory mucosa
- Can transmit viral infections
Preventing Electrothermal Damage #
- Always keep the active electrode in full view.
- Tenting Technique: Lift tissue away from surrounding structures before activating the ESU.
- When to Activate:
- Only activate the ESU when you have a clear view of the field and the electrode is in contact with the tissue. Avoid non-contact activation.
- Power Usage: Use the lowest voltage mode necessary to achieve the desired effect. Avoid overshooting power settings.
- Handling the Electrode:
- The tip stays hot for several secondsβdon’t let it touch anything inside or outside the body. The scrub nurse should place it in a quiver.
- Single Use Electrodes:
- Discard after one use; do not reuse.
- Consider Alternatives:
- Use bipolar or ultrasonic electrosurgical devices if available.